E Commute
No.
Name
Description
E01
NC/NO
Sets whether the OCT output is open or closed when the condition is not met. The default setting is open.
(Only Applicable to Model FP-AS310A)
E02
OCT Info
Sets the definition for the flow meter’s OCT operation. Default: Off. (Only Applicable to Model FP-AS310A)
E03
Pulse Width
Sets the pulse width of the flow meter when E02 OCT Definition is set to “Pulse v2”. Default: 0.5 ms.
(Only Applicable to Model FP-AS310A)
E04
Modbus ID
Set up Modbus ID location, default 1.
(Only applicable to Model FP-AS310R)
E05
Baud
Sets the Modbus communication baud rate for the flow meter. The Master and Slave must use the same baud rate for communication. Default: Off. (Only applicable to Model FP-AS310R)
E06
Iout Def.(Iout Definition)
Sets the data to be output via the analog signal. The unit follows the setting in Parameter B (Unit Configuration). Default: Off. (Only Applicable to Model FP-AS310A)
E07
4mA Value
Set up corresponding value to 4mA, default 0. Both positive and negative values work. The value should be smaller than 20mA.
(Only Applicable to Model FP-AS310A)
E08
20mA Value
Set up corresponding value to 20mA, default 100. Both positive and negative values work. The value should be larger than 4mA.
(Only Applicable to Model FP-AS310A)
E11
4mA Adj. uA(4mA Output Adjustment in microAmp)
Fine-tunes the flow meter’s 4 mA output value. Unit: µA (microamperes). Default: 0.
(Only Applicable to Model FP-AS310A)
E12
20mA Adj. uA(20mA Output Adjustment in microAmp)
Fine-tunes the flow meter’s 20 mA output value. Unit: µA (microamperes). Default: 0. (Only Applicable to Model FP-AS310A)
E14
PulseV2 Unit
Defines the flow rate corresponding to one pulse output from the flow meter.
(Only Applicable to Model FP-AS310A)
E15
Byte format
Sets the data format for Modbus RTU communication. Default: N,8,1 (No parity, 8 data bits, 1 stop bit).
(Only applicable to Model FP-AS310R)
E16
Bit order
Sets the bit order for each byte during Modbus RTU transmission.
(Only applicable to Model FP-AS310R)
E17
Endianness
Sets the byte order for multi-byte data during RX (flow meter response) transmission.
(Only applicable to Model FP-AS310R)