D06 Low LIM

Definition

Set the lower limit of the normal flow range. Default: 0.

(D04 , D05 , D06 , and D07 work together to enable the flow range over-limit alarm function.)

Flow Range Limit Alarm Example (Click to Expand)
D05 Upper Limit: 30 LPM
D06 Lower Limit: 10 LPM
D07 Hysteresis: 3% (default)

Calculations:
Range: 30 − 10 = 20 LPM
Hysteresis: 20 × 3% = 0.6 LPM

Alarm Behavior:
When flow exceeds the upper limit (30 LPM), the alarm is triggered.
The alarm will clear only when the flow drops below 29.4 LPM (30 − 0.6).
When flow falls below the lower limit (10 LPM), the alarm is triggered.
The alarm will clear only when the flow rises above 10.6 LPM (10 + 0.6).

Notes

If either D05 or D06 is adjusted, the other must also be set; otherwise, the display may not function correctly. If both are unset, default values are D05 = 100, D06 = 0.

DND Mode

When Do Not Disturb (DND) mode is enabled:

  • The flow meter detects abnormal conditions but the red backlight will not turn on. Example: Alarm-A “Paddle wheel abnormality” will not trigger the backlight in DND mode.

  • Abnormal conditions are still displayed on the Status/Alarm page.

  • If connected via RS485, the monitoring system still receives the alarm signal.

  • OCT alarm outputs continue to operate if configured.

Operation Procedure

Button Functions (Click to Expand)

The main unit has four buttons. The bottom line (Line 3) of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:

  • Blue Button: Leftmost button: Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button: Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses

Flow Range Over-Limit Alarm Operation (Click to Expand)

When the flow range alarm is enabled and the flow reaches the alarm condition, Alarm B is triggered. Alarm bits (Bit0, Bit1) display 1 to indicate the alarm has been activated.

The system responds according to the alarm mode set in D04:

  • On: The receiving device receives the alarm signal, and the flow meter screen displays a red backlight warning.

  • Do Not Disturb (DND): The receiving device still receives the alarm signal, but the screen does not display the red backlight.

1

Set D04 to enable the flow range over-limit alarm

2

Set D05 to define the flow upper limit

3

Set D06 to define the flow lower limit

4

Set D07 to define the flow hysteresis

5

Flow Range Over-Limit Alarm Operation

When the flow range alarm is enabled and the flow reaches the alarm condition, Alarm B is triggered. Alarm bits (Bit0, Bit1) display 1 to indicate the alarm has been activated.

The system responds according to the alarm mode set in D05:

  • On: The receiving device receives the alarm signal, and the flow meter screen displays a red backlight warning.

  • Do Not Disturb (DND): The receiving device still receives the alarm signal, but the screen does not display the red backlight.

D06 Options, RS485 Values, Read/Write Commands, and Formats

Parameter Input Range
Corresponding Values
RS-485 Value

Default

0

-

Minimum

-100000

-

Maximum

99999

-

Last updated