B05 Dec. Place

Definition

Sets the number of decimal places displayed on the LCD and LED (up to 2 places). Values are truncated (rounded down). Default: 02 (2 decimal place).

Notes

This parameter does not affect calculations; it only affects the display. Cumulative flow is still accumulated using the original data.

Operation Procedure

Button Functions (Click to Expand)

The main unit has four buttons. The bottom line (Line 3) of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:

  • Blue Button: Leftmost button: Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button: Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses

B05 Dec. Place Setting Steps
• From the main screen (green backlight), 
  press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select B.Unit , press Orange Button (Enter).
• Use Gray Arrow Buttons to select B05 Dec. Place, press Orange Button (Enter).
• Use the arrow Buttons to set the number of decimal places:
  Left Gray Button: increase value / Right Gray Button: decrease value
• Press Orange Button (Set) to save.
• Press Blue Button (ESC) to return to the main screen (green backlight).

B05 Dec. Place (Decimal Place)

Specifies the number of decimal places displayed on the LCD screen. Data is truncated without rounding.

  • Maximum of 4 decimal places, default is 4.

  • This parameter affects display only, not calculations; cumulative flow uses original data.

Parameter Input Range
Values

Default

1

Minimum

0

Maximum

2