A Measurement Setup

A Measure parameter list

No.

Name

Description

A01

Pipe Type

Set the T-connector spec to use the correct calibration data. The default is 1 inch.

A02

Scales Factor

Sets the correction method for instantaneous flow by fine tuning flow volume in proportion. Default: 1.0

A04

Damping

Smooths flow reading by decreasing the fluctuation of measurement data, the lowest is 0, default value is 3 seconds.

A05

Direction

Set the displayed flow direction. Default: normal flow.

A09

Manual Offset

Corrects fixed errors that do not vary with flow (zero offset)