Parameters

Parameter categorizing table

Parameter Group

Name

Description

A

Measure

A group is about flow volume measurement

B

Unit

B group is about unit

C

Total & History

C group is about accumulated and history record

D

Alarm & Info(Alarm & Information)

D group is about alarm and notification.

E

Commute

E group adjust NPN 4-20mA and RS485

F

Simulate

F group is about simulation mode for communication and alarm testing.

H

System

H group is for system parameters, like brightness and reset.

Parameter description

A Measure parameter list

No.

Name

Description

A01

Pipe Type

Set the T-connector spec to use the correct calibration data. The default is 1 inch.

A02

Scales Factor

Sets the correction method for instantaneous flow by fine tuning flow volume in proportion. Default: 1.0

A04

Damping

Smooths flow reading by decreasing the fluctuation of measurement data, the lowest is 0, default value is 3 seconds.

A05

Direction

Set the displayed flow direction. Default: normal flow.

A09

Manual Offset

Corrects fixed errors that do not vary with flow (zero offset)

B Unit parameter list

No.

Name

Description

B01

Vol. Unit(Volume Unit)

Sets the flowmeter’s instantaneous flow volume unit. Default: L (liter).

B02

Time Unit

Sets the time unit for instantaneous flow. Default: minute.

B03

Total Unit(Total Volume Unit)

Sets the volume unit for cumulative and historical flow. Default: L (liter).

B05

Decimal Places

Sets the number of decimal places displayed on the LCD by using round down method. The maximum number of digits is 4 digits.Default: 04 (4 decimal place).Cumulative flow volume still calculated by using original value.

C Total & History parameter list

No.

Name

Description

C02

Del. Total (Delete Total)

Set the flow meter’s totalizer to zero manually. Default: No.

D Alarm & Information parameter list

No.

Name

Description

D01

Paddle Ala.(Paddle Alarm)

Set the flow meter’s alarm mode when module detect pipe diameter is abnormal. Default: Do Not Disturb. The alarm code belongs to Alarm – A.

D04

FlowLIM Ala (Flow Volume Limit Alarm)

Set the flow meter’s alarm mode for flow exceeding the instantaneous flow range. Default: Off. The alarm code belongs to Alarm – B.

D05

Up LIM(Upper Limit)

Set the upper limit of the normal flow range. Default: 100.

D06

Low LIM(Lower Limit)

Set the lower limit of the normal flow range. Default: 0.

D07

Hysteresis% (Hysteresis Percentage)

Set the hysteresis percentage for clearing flow upper and lower limit alarms. This ensures the alarm is only cleared when the flow stabilizes back within the set range. The percentage is defined relative to the span between the upper and lower limits. Default: 03 (3%).

D08

Total Ala.(Total Alarm)

Set the flow meter’s alarm mode for totalizer values exceeding the set range. Default: Off. The alarm code belongs to Alarm – B. (D08 , D9 , and D10 work together to enable totalizer target alarm functionality.)

D09

Total Trig.(Total Alarm Trigger Condition)

Set the notification condition for the totalizer target. Default: Greater than target.

D10

Total Target

Set up alarm target value of cumulative flow speed. Default: +100.

D13

PulseOverlap

Set the flow meter’s alarm mode for overlapping output signals caused by an excessively long E03 Pulse Width. The alarm code belongs to Alarm – B. (Only Applicable to Model FP-AS310A)

D14

IoutStatAlarm

Set the flow meter’s alarm mode for abnormal 4–20 mA current output (Iout). Default: Do Not Disturb. The alarm code belongs to Alarm – C. (Only Applicable to Model FP-AS310A)

E Commute parameter list

No.

Name

Description

E01

NC/NO

Sets whether the OCT output is open or closed when the condition is not met. The default setting is open.

(Only Applicable to Model FP-AS310A)

E02

OCT Info

Sets the definition for the flow meter’s OCT operation. Default: Off. (Only Applicable to Model FP-AS310A)

E03

Pulse Width

Sets the pulse width of the flow meter when E02 OCT Definition is set to “Pulse v2”. Default: 0.5 ms.

(Only Applicable to Model FP-AS310A)

E04

Modbus ID

Set up Modbus ID location, default 1.

(Only applicable to Model FP-AS310R)

E05

Baud

Sets the Modbus communication baud rate for the flow meter. The Master and Slave must use the same baud rate for communication. Default: Off. (Only applicable to Model FP-AS310R)

E06

Iout Def.(Iout Definition)

Sets the data to be output via the analog signal. The unit follows the setting in Parameter B (Unit Configuration). Default: Off. (Only Applicable to Model FP-AS310A)

E07

4mA Value

Set up corresponding value to 4mA, default 0. Both positive and negative values work. The value should be smaller than 20mA.

(Only Applicable to Model FP-AS310A)

E08

20mA Value

Set up corresponding value to 20mA, default 100. Both positive and negative values work. The value should be larger than 4mA.

(Only Applicable to Model FP-AS310A)

E11

4mA Adj. uA(4mA Output Adjustment in microAmp)

Fine-tunes the flow meter’s 4 mA output value. Unit: µA (microamperes). Default: 0.

(Only Applicable to Model FP-AS310A)

E12

20mA Adj. uA(20mA Output Adjustment in microAmp)

Fine-tunes the flow meter’s 20 mA output value. Unit: µA (microamperes). Default: 0. (Only Applicable to Model FP-AS310A)

E14

PulseV2 Unit

Defines the flow rate corresponding to one pulse output from the flow meter.

(Only Applicable to Model FP-AS310A)

E15

Byte format

Sets the data format for Modbus RTU communication. Default: N,8,1 (No parity, 8 data bits, 1 stop bit).

(Only applicable to Model FP-AS310R)

E16

Bit order

Sets the bit order for each byte during Modbus RTU transmission.

(Only applicable to Model FP-AS310R)

E17

Endianness

Sets the byte order for multi-byte data during RX (flow meter response) transmission.

(Only applicable to Model FP-AS310R)

F Simulate parameter list

No.

Name

Description

F01

Sim. Volume (Simulation Volume)

Sets the flow meter’s simulated instantaneous flow. The unit follows the setting in Category B. Default: +60.

F02

Sim. Mode (Simulation Mode)

Enables or disables the flow meter’s simulation mode. In simulation mode, the meter displays user-defined virtual flow values instead of actual sensor measurements. Default: Off.

F03

Sim. PulseHz (Simulation PulseHz)

Sets the simulated OCT pulse output frequency, in pulses per second (Hz). Range: 0–700. Default: 0.

(Applicable to Model FP-AS310A))

H System parameter list

No.

Name

Description

H02

LCD Bri.(LCD Brightness)

Set up brightness of LCD with 3 levels, default middle level.

H03

Green Light

Set up time of LCD green lighting with 3 durations, 10s, 60s, and always on.

H04

Language

Set up language. There are 2 languages:English, Traditional Chinese, default off.

H05

Paddle Init.(Paddle Initialization)

Press "Start" to initialize the paddle. User needs to rotate the paddle at least 20 rounds until the screen shows "Init. Succeed". If screens shows "Init. Failed", please try again. If the problem persists, please replace your paddle.

H06

Reset

Clear all the data (include date and cumulative data), reset all parameters to default values.

H07

Lock Device

Turn on the password function to prevent unauthorized access to the setting screen. The password can be changed in parameter H08, default off.

H08

Password

Modify the 4 digits password to prevent unauthorized access to the setting screen. The password function is enabled at parameter H07. The preset password is 0000.

H09

Engineer mode

H10

System Info

H11

Home Page