Error Type
Causes
Solutions
Bit 0
ToF measurement timeout
Initial setting error
Probe not connected or poorly set up
Initial setting error
Check probe installation
Bit 1
No signal
Probe not installed or poorly installed
Couplant is dry
Too much grime inside the pipe
Pipe liner exists but not set
1.Check probe installation
2. Re-coat coupling agent
3. Clean up the pipeline
4. Check initial settings
Bit 2
Thermal sensor short circuit
Sensor short circuit
Instrument failure
Check wiring
Contact manufacturer
Bit 3
Thermal sensor open circuit (disconnected)
Sensor not installed
Sensor disconnected
Instrument failure
Check wiring
Check wiring
Contact manufacturer
Bit 4
Empty pipe
Probe not connected or poorly set up
No fluid in the pipe
1.Check probe installation 2. Confirm pipe fluid status
Bit 5
Poor signal quality
Probe installed incorrectly
Too much grime in the pipe
Bubbles in the pipe
Bubbles in the pipe
Check probe installation
Clean pipe
Install probe on side surface or change installation setup location
Change installation setup location
Bit 6
Incorrect initial value settings
Initial value setting error
Recheck initial value settings
Bit 7
Network communic ation error
Communication module failure
Reboot
Contact manufacturer