F Category Options and Descriptions
F01 Sim. Rate (Simulation Rate) - Simulated Flow Defines the instantaneous flow rate to simulate. The unit applied will follow the settings defined in the B classification parameters.
Data Type
float32
Default
60
Minimum
-100000
Maximum
100000
F02 Sim. Mode (Simulation Mode) Activating this option enables the simulation function, which is disabled by default. When enabled, the OCT outputs (excluding pulse), RS485 communication, and analogue current output will operate based on the simulated data and changes in the totalised flow.
Please note that the totalised flow and historical data will be affected.
The main screen will display an indication that the system is in simulation mode. Ensure the relevant communication functions are enabled for proper operation.
OptionsRS-485 NumbersRemarksOFF
0
Default
ON
1
After selecting a command, confirm whether to Save or Not Save.
Last updated